Climatic environment testing - UV ultraviolet test

Test the project

UV test

 

Reference standards

GB/T16422.3-2014; ISO 4892-3-2016

 

Range of competence

 UVA340/UVB313/UVC254

 

UV aging belongs to sunlight irradiation aging, which is usually used to evaluate the rate of product aging under ultraviolet radiation. Sunlight irradiation aging is the main aging damage suffered by outdoor materials, and for indoor materials, they will also be subjected to a certain degree of sunlight irradiation aging or aging caused by ultraviolet radiation in artificial light sources (such as ultraviolet lamps, ultraviolet bands in fluorescent lamps, etc.).

 

UV lamp aging test is a test method that simulates the ultraviolet band of sunlight using artificial light sources to verify the environmental aging resistance of polymer materials such as organic coatings, plastics, and rubber. The UV aging test has three modes: illumination, condensation, and spraying. The lighting stage is used to simulate daytime in natural environments, and the lighting intensity and temperature can be controlled during the lighting period; The condensation stage is used to simulate the phenomenon of condensation on the surface of the sample at night. During the condensation stage, the UV lamp is turned off (in a dark state) and only the test temperature is controlled; The spraying stage simulates a rainy environment by continuously spraying water onto the surface of the sample.

 

Types and Applications of UV Aging Lamp Tubes
There are three commonly used types of lamps for UV aging, namely UVA-340, UVA-351, and UVB-313. All types of lamps emit ultraviolet light instead of infrared or visible light. Due to the different total energy and spectral distribution emitted by these lamps, the specific type of lamp used 

should depend on the application conditions.

 

Long wave ultraviolet lamps (UVA lamps) do not damage materials as quickly as short wave ultraviolet rays, but they are closer to real outdoor aging. There are two common types of lamps: UVA-340 and UVA-351. The spectral distribution of UVA-340 lamps is similar to the ultraviolet band of outdoor sunlight, and is generally used for light aging testing of conventional outdoor products. UVA-351 lamps can simulate the ultraviolet part of sunlight passing through glass windows, mainly used for indoor product aging.

Short wave ultraviolet lamps emit non natural short wave ultraviolet radiation with shorter wavelengths, causing materials to corrode faster. Common examples include UVB-313 lamps. The conditions for UVB-313 lamps are the most stringent, as they can generate high-intensity ultraviolet radiation and cause material aging faster than UVA lamps, thereby achieving the goal of accelerating aging as quickly as possible. They are particularly useful for testing very durable materials.

 

Lamp UVA-340: simulates the ultraviolet radiation of sunlight and is commonly used for light aging testing of outdoor products.

Lamp UVB-313: Used to accelerate material aging and study material durability, it is a relatively fast and time-saving photo aging test.

Lamp UVA-351: Used for indoor environmental UV aging testing, simulating the solar spectrum of sunlight filtered through window glass.

UV aging test conditions: lamp type, irradiation intensity, temperature, time, cycle times, non light type (condensation, spray) conditions, etc.

Applicable product field scope
Rubber, plastics and their products, automotive materials, coatings, paints, packaging materials, building materials, textiles, roofing materials, etc.

     