What are the factors that affect BCI test results?

Author.

LCS

Source:

Post time:

2026-10-09

In automotive electronic EMC testing, BCI (high current injection) is an important immunity test. It couples RF interference to the tested component harness through a current injection probe to investigate whether automotive electronic components can function properly in RF interference environments.


 

But in practical work, many laboratories have encountered such situations:

Sometimes there may be differences in the test results of the same product in different laboratories, or differences in the same laboratory due to different testing setups. What exactly is the reason for this? Is there a calibration issue? Is there a difference in equipment between the two laboratories?


 

The BCI test results are influenced by multiple factors. This article explores several key factors that affect BCI test results from the perspective of practical testing.

 

1、 Placement details: centimeter level deviation, resulting in significant differences

BCI testing essentially couples high-frequency electromagnetic energy to the product's wiring harness by injecting clamps. The wiring harness here is like a 'receiving antenna', and when the antenna state changes, the energy coupled in will undergo a huge change.


 

Height of wire harness above ground (distributed capacitance)

The standard usually requires that the harness be 50mm away from the metal ground. If the height of the foam block used for support deviates by 1cm, or the middle of the harness sags slightly, the distributed capacitance between the harness and the ground will change. In the high frequency range, this small change in capacitance is enough to cause a difference of several dB in coupling efficiency.


 

Injection position of probe

The distance between the injection probe and the EUT connector (e.g. 150mm, 450mm) is strictly specified in the standard. High frequency signals will form standing waves (peaks and valleys) on the line beam. The probe is slightly moved a few centimeters, and it may have just moved from the "standing wave node (minimum energy)" to the "standing wave belly point (maximum energy)", instantly doubling the interference that the product can withstand.


 

Disposal of excess wiring harness

Is the wiring harness straightened, laid flat, or coiled? The number of turns and folding method of the winding will directly change the common mode impedance of the wire harness.


 

2、 Laboratory environment: invisible 'impedance differences'

From the appearance, the shielding rooms are identical, but in fact, the "bottom line" of each laboratory is different.


 

Reference ground plane and ground impedance

The current injected by BCI will eventually flow back to the amplifier through the grounding circuit. The material, degree of oxidation, impedance of the board joint, and impedance of the grounding cable of the laboratory grounding plane will all change the resonance point of the entire loop.


 

3、 Calibration and Control Algorithm: 50 Ω Fixture vs Complex Actual Wiring Harness

Before the official start of BCI testing, a 50 Ω standard calibration fixture is required to calibrate and calculate the required forward power injection. But the problem lies here:


 

Mismatch of harness impedance

The calibration fixture is a standard 50 Ω system, but the common mode impedance of the wiring harness in actual cars or equipment may range from a few ohms to several thousand ohms and fluctuate dramatically with frequency.


 

Power Limitation in Closed Loop Method

At present, closed-loop testing uses a system to monitor the actual current on the wiring harness, and if it is not enough, the injection power is increased. When the wire harness exhibits high impedance, the system will continue to apply power until it reaches the "maximum injection power" limit in order to achieve the target current.


 

Small deviations in the calibration of this power limit by different laboratories, or differences in the response speed of system probes, can result in differences in the actual energy injected into the product.


 

4、 Auxiliary equipment (AE) and power supply: the overlooked 'other end'

One end of the test harness is connected to the product (EUT), and the other end is connected to the auxiliary equipment (AE).


 

Filtering and grounding at AE end

If the shielding box at the AE end changes, the capacitor of the filter board decays, or the grounding resistance of the AE itself is different after switching to another laboratory, it will cause a complete change in the impedance of the wiring harness terminal.


 

Artificial power network

Although LISN provides standard power impedance, the routing of power lines and the noise ripple of the power supply itself can still modulate sensitive frequency points.


 

5、 How to improve the consistency of BCI testing

 

  Strictly fix the wiring harness placement: use standard insulation brackets, strictly control the height of the wiring harness above the ground, record the precise position of the probe and take photos for archiving.

Fixed auxiliary testing environment: During rectification and certification testing, try to use the same set of AE, testing harness, and load as much as possible.

Verify the calibration coefficient of the monitoring clamp, and rely on the monitoring clamp to read the current for closed-loop testing. If there is a deviation in the calibration file of the laboratory probe, it can cause the system to "misjudge" the magnitude of the current, resulting in excessive or insufficient interference.


 

BCI testing is never just about "clamping the probe and starting". By understanding these micro variables, the next time we encounter inconsistent laboratory results, we will know where to look for the cause.


 